+886-2-8772-8910
Parameter |
L9455 series |
L9456 series |
Unit |
|
---|---|---|---|---|
Arc size |
1.5 |
3 |
mm |
|
Window material |
UV glass |
— |
||
Emission wavelength range |
185 to 2000 |
nm |
||
Main discharge voltage variable range *1 |
400 to 600 |
V |
||
Main discharge capacitor *2 |
0.22 / 0.11 / 0.047/ 0.28 |
µF |
||
Maximum input energy (per flash) *3 |
See operating conditions. |
— |
||
Maximum average input (continuous) *4 |
5 |
W |
||
Light output stability (Max.) *5 |
2 |
1.5 |
% CV |
|
Guaranteed life *6 |
1×109 |
flashes |
||
Input voltage range |
11 to 28 |
V |
||
Input current |
1 |
A |
||
Inrush current |
4 |
A |
||
Trigger input |
Rectangular wave 5 V to 10 V; |
|
||
Trigger input impedance |
330 |
Ω |
||
Cooling method |
Not required *8 |
— |
||
Weight (standard type: L9455-01) |
Approx. 155 |
g |
||
Operating temperature range |
0 to +40 |
°C |
||
Storage temperature range |
-40 to +90 |
°C |
||
Storage humidity range |
Below 95 (no condensation) |
% RH |
||
EMC standards |
Without EMC noise filter |
IEC61326-1: 2005 Group1, ClassA IEC61326-1: 2005 Group1, ClassB |
— |
|
With EMC noise filter |
||||
Safety standards |
IEC62471: 2006 Risk Group3 |
|
Type No. |
Main discharge |
Main discharge |
Maximum input |
Maximum |
Maximum input |
---|---|---|---|---|---|
L9455/L9456-○1 |
0.22 |
400 |
17.6 |
284 |
5.0 |
500 |
27.5 |
182 |
5.0 |
||
600 |
39.6 |
126 |
5.0 |
||
L9455/L9456-○2 |
0.11 |
400 |
8.8 |
530 |
4.7 |
L9455/L9456-○2 |
500 |
13.8 |
362 |
5.0 |
|
L9455/L9456-○3 |
0.047 |
400 |
3.8 |
530 |
2.0 |
500 |
5.9 |
530 |
3.1 |
||
600 |
8.5 |
530 |
4.5 |
||
L9455/L9456-○4 |
0.28 |
400 |
22.4 |
223 |
5.0 |
500 |
35.0 |
142 |
5.0 |
||
600 |
50.4 |
100 |
5.0 |
Parameter |
L11316 series |
L11317 series |
Unit |
||
---|---|---|---|---|---|
Arc size |
1.5 |
3 |
mm |
||
Window material |
UV glass |
— |
|||
Emission wavelength range |
185 to 2000 |
nm |
|||
Main discharge voltage |
Internal |
650 to 1000 |
V |
||
Main discharge voltage variable range① |
External |
500 to 1000 |
|||
Main discharge capacitance② |
0.2 /0.1 |
µF |
|||
Maximum input energy (per flash)③ |
See operating conditions. |
J |
|||
Maximum average input (continuous)④ |
5 |
W |
|||
Light output stability (Max.)⑤ |
3 |
2.5 |
% CV |
||
Guaranteed life⑥ |
5 × 108 |
flashes |
|||
Input voltage range |
21.6 to 26.4 |
V |
|||
Input current |
0.75 |
A |
|||
Inrush current |
3 |
A |
|||
Trigger input |
Rectangular waveform 5 V to 10 V; |
— |
|||
Trigger input impedance |
330 |
Ω |
|||
Cooling method |
Not required⑧ |
— |
|||
Weight (standard type: L11316-01) |
Approx. 328 |
g |
|||
Operating temperature range |
0 to +40 |
°C |
|||
Storage temperature range |
-40 to +90 |
°C |
|||
Storage humidity range |
Below 95 (no condensation) |
% RH |
|||
EMC standards |
Without EMC noise filter |
IEC61326-1: 2005 Group1, ClassA |
— |
||
With EMC noise filter |
IEC61326-1: 2005 Group1, ClassB |
||||
Safety standards |
IEC61010-1: 2010 |
— |
|||
IEC62471: 2006 Risk Group3 |
①Internal: Adjustable with variable trimmer.
External: Variable with control voltage from 2.44 V to 4.88 V.
②Installed at time of shipment (refer to the above lineups).
③Maximum lamp input energy (per flash)
E=1/2 CV2
E: Maximum lamp input energy (J) V: Main discharge voltage (V) C: Main discharge capacitance (F)
④Maximum average lamp input (continuous)
W=E × f
f: Lamp emission repetition frequency (Hz)
⑤Light output stability is given by: Light output stability (% CV) = light output standard deviation / average light output × 100
⑥At 5 W operation
⑦Only for external control;synchronized with rising edge.
⑧Cooling is required when the package temperature exceeds 50 °C during operation.
Type No. |
Main discharge capacitor |
Main discharge voltage |
Maximum input energy |
Maximum repetition rate |
Maximum input |
---|---|---|---|---|---|
L11316- ○1 |
0.2 |
500 |
25 |
200 |
5 |
700 |
49 |
102 |
5 |
||
1000 |
100 |
50 |
5 |
||
L11316- ○2 |
0.1 |
500 |
12.5 |
400 |
5 |
700 |
24.5 |
204 |
5 |
||
1000 |
50 |
100 |
5 |